qPCR: poor efficiency
Standard-curve efficiency outside the lab’s accepted window.
Ranked causes
- Inhibitors in template or standards (carry-over from prep)Efficiency improves after dilution or clean-up of the same samples.
- Standard dilution series inaccurate (pipetting / diluent)R² poor or points zigzag; remade dilutions straighten the curve.
- Primer concentration or amplicon design suboptimalSame standards give good efficiency with a different validated assay.
- Baseline/threshold settings distort early or late cyclesManual threshold in exponential phase restores efficiency near 100%.
Recovery steps
- Remake the standard curve with fresh serial dilutions (new tips each step)Dilution error is the most common efficiency failure.
- Dilute samples/standards further to outrun inhibitors; compare efficiencyInhibition presents as efficiency << 100% that recovers on dilution.
- Re-analyse with threshold set in the exponential phase for all wellsAnalysis artefacts can mimic chemistry problems.
Controls first
- Is the standard-curve R² also poor, or only the efficiency slope off?If yes: Dilution series or pipetting — remake standards before redesigning primers.
Related symptoms
Opens Diagnostics capture with this first-party playbook behind it. Sign in if the workspace asks.