qPCR: multiple melt peaks
Melt analysis shows more than one product peak for a supposed single-amplicon assay.
Ranked causes
- Non-specific amplification from primers (anneal too low / design)Extra peak shrinks when anneal is raised 2–3 °C.
- Primer-dimer peak at low Tm alongside specific productLow-Tm peak present in NTC as well as samples.
- Contaminating template producing a second ampliconSecond peak matches a known off-target size on gel; fresh reagents help.
- Heterozygous indel / isoform mix in the sample biologyTwo specific-Tm peaks stable across clean NTCs and replicate preps.
Recovery steps
- Raise annealing temperature 2–3 °C and re-check meltOften removes non-specific peaks without losing the true product.
- Run NTC melt; if low-Tm peak matches samples, redesign or lower primer concentrationPrimer-dimer is diagnosed when NTC shares the extra peak.
- Spot-check on a gel or switch to a hydrolysis-probe chemistryConfirms whether extra fluorescence is a real second amplicon.
Controls first
- Does the NTC also show one of the extra melt peaks?If yes: Primer-dimer or contamination — not sample isoform biology.