qPCR: flat amplification curve
Fluorescence stays at baseline across cycles despite expected template.
Ranked causes
- Probe or SYBR omitted / degraded in the master mixGel of the same reaction shows product; fluorescence channel stays flat.
- No template or failed RT (empty input)Positive-control well curves normally; sample wells stay flat.
- Wrong filter / reporter channel selected on the instrumentRaw data in another channel shows the curve; assay chemistry is intact.
- Cycling program never reaches extension / read stepProtocol file missing plate-read; fixing the program restores curves.
Recovery steps
- Verify reporter channel / filter matches dye (e.g. FAM vs HEX) in the run setupWrong channel is a frequent flat-curve cause after copying a template run.
- Re-make mix with fresh dye/probe; include a kit positive controlSeparates detection chemistry from template problems.
- Confirm the cycling file includes a plate-read at the end of extensionMissing read steps produce empty datasets that look like chemistry failure.
Controls first
- Did the positive-control well also stay flat in the selected channel?If yes: Channel, dye/probe, or cycling program — not a single sample.