qPCR: high replicate ct cv
Technical replicate Cts disagree beyond the lab’s usual spread.
Ranked causes
- Master mix incompletely mixed before dispenseCV increases toward later-plated wells; vortexing before dispense helps.
- Baseline drift on a subset of wells (analysis)Raw curves look similar; Ct disagreement shrinks with shared baseline settings.
- Inconsistent pipetting of master mix or templateCV worse on manually plated wells than on a calibrated dispenser.
- Evaporation / edge effects on the plate during the runOutlier wells sit on the plate perimeter; optical seal incomplete.
Recovery steps
- Re-plate with reverse pipetting; discard bubbled wellsVolume inconsistency is the top driver of technical Ct CV.
- Vortex master mix immediately before every dispense roundPrevents settling across a long plating session.
- Replace optical seal; avoid leaving edge wells empty without buffer moatReduces evaporation-driven perimeter outliers.
Controls first
- Are replicate CVs high on standard-curve / positive-control wells as well?If yes: Pipetting, mix, evaporation, or analysis — not sample biology alone.