Assay / ELISA / Binding: poor standard curve
Standard curve R² / shape unfit for quantification (non-linear, noisy, or inverted).
Ranked causes
- Degraded or wrongly reconstituted standard stock (lot / storage)Fresh standard lot recovers the curve; samples unchanged.
- Standard dilution series pipetting errorPoints zigzag; remade dilutions restore a smooth sigmoid/linear segment.
- Incubation time/temperature drifted during standard platingEarly vs late standards diverge; timed incubations fix shape.
- Reader wavelength or blanking mis-set for the substrateCorrect λ restores expected OD span on the same plate.
Recovery steps
- Remake standards with fresh serial dilutions and new tips each stepPipetting error is the most common poor-curve cause.
- Reconstitute a new standard vial per kit instructions; avoid extra freeze-thawDegraded stock cannot be rescued by dilution technique alone.
- Verify reader wavelength and blank against kit substrate instructionsWrong λ compresses or warps the apparent curve.
Controls first
- Do positive-control / kit-standard wells also fail to form a usable standard curve?If yes: Standards, incubation, or reader setup — not sample unknowns alone.