Assay / ELISA / Binding: saturated curve
Signal plateaus across a wide concentration range instead of a sigmoidal response.
Ranked causes
- Sample concentration too high — off the top of the standard curveMultiple sample dilutions all read at the assay's OD ceiling.
- Detector conjugate lot concentration too high, driving max signal earlyCurve compresses toward saturation at all but the lowest standards.
- Substrate development run to completion (all-or-none color)Standards also plateau at the same OD.
Recovery steps
- Serially dilute samples 1:5 to 1:50 and re-readBrings concentrated samples back onto the linear/sigmoidal portion of the curve.
- Read the plate earlier, before substrate fully developsPrevents standards themselves from plateauing at OD ceiling.
- Titrate detection conjugate down 2-foldWidens the assay's dynamic range before saturation.
Controls first
- Do the standard-curve / positive-control wells also plateau at the top OD?If yes: Assay-level saturation (substrate/detector), not just sample concentration.
- Do the standard-curve / positive-control wells also plateau at the top OD?If no: Sample-specific — dilute and re-read.