Assay / ELISA / Binding: no signal
No readable OD/fluorescence above blank, including expected positive wells.
Ranked causes
- Detection substrate omitted, expired, or wrong for the enzymeFresh substrate on a known-positive well develops immediately.
- Capture/detection antibody omitted or wrong orderChecklist shows a skipped incubate; re-running with antibodies restores signal.
- Reader set to wrong mode/wavelength (flat zeros)Same plate reads normally after correcting λ / luminescence mode.
- Plate never coated / coating washed off before assayRe-coat overnight restores kit positive control.
Recovery steps
- Apply fresh substrate to a known-positive well and watch for developmentFastest split between substrate vs upstream assay failure.
- Confirm reader mode/wavelength matches the kitWrong readout mode presents as total zeros.
- Re-coat and re-run with a printed step checklist for antibody orderSkipped coating/antibody steps are common total-loss causes.
Controls first
- Did kit positive-control / top standard wells also read at blank level?If yes: Substrate, antibodies, coating, or reader — not only unknowns.
Related symptoms