Assay / ELISA / Binding: no signal

No readable OD/fluorescence above blank, including expected positive wells.

  1. Ranked causes

    1. Detection substrate omitted, expired, or wrong for the enzymeFresh substrate on a known-positive well develops immediately.
    2. Capture/detection antibody omitted or wrong orderChecklist shows a skipped incubate; re-running with antibodies restores signal.
    3. Reader set to wrong mode/wavelength (flat zeros)Same plate reads normally after correcting λ / luminescence mode.
    4. Plate never coated / coating washed off before assayRe-coat overnight restores kit positive control.
  2. Recovery steps

    1. Apply fresh substrate to a known-positive well and watch for developmentFastest split between substrate vs upstream assay failure.
    2. Confirm reader mode/wavelength matches the kitWrong readout mode presents as total zeros.
    3. Re-coat and re-run with a printed step checklist for antibody orderSkipped coating/antibody steps are common total-loss causes.
  3. Controls first

    • Did kit positive-control / top standard wells also read at blank level?If yes: Substrate, antibodies, coating, or reader — not only unknowns.