Assay / ELISA / Binding: hook effect
Very high analyte reads paradoxically low until diluted (prozone/hook).
Ranked causes
- Analyte concentration above the assay’s hook threshold1:10–1:100 dilution raises apparent concentration into the standard range.
- Antibody pair affinity/concentration set too low for high-load samplesRicher detection antibody reduces hooking on the same samples.
- Insufficient wash leaving unbound analyte that blocks detectionExtra washes plus dilution both move results upward.
- Standard curve too narrow to flag the hook visuallyExtending standards higher reveals non-monotonic sample behaviour.
Recovery steps
- Re-assay samples at 1:10 and 1:100; report the dilution that falls on-curveDilution test is the definitive hook check.
- Increase detection antibody concentration one step if dilutions still hookSome kits need richer detection for high-load matrices.
- Extend the standard curve upward for future runs of this sample typeMakes hooking visible before results are reported.
Controls first
- Do standard-curve / positive-control wells look normal while neat high samples read paradoxically low?If yes: Hook/prozone behaviour — dilute samples and re-report; standards are not the failure.
- Do diluted samples read higher (on-curve) than the neat sample?If yes: Confirmed hook — report the on-curve dilution.