Western blot: uneven transfer
Patchy or gradient protein transfer across lanes or the membrane.
Ranked causes
- Air bubbles or incomplete sandwich contact during transferCircular voids on Ponceau align with sandwich centre or roller gaps.
- Transfer buffer pH / methanol composition off for the membrane typeGradient or edge-weak transfer after a fresh buffer prep; pH drifts from 8.3.
- Transfer time or voltage too short for large proteins (or too aggressive for small)High-MW targets retained in gel; low-MW targets blow through to filter paper.
- Power supply current drift or loose electrode contactCurrent log unstable; other tanks with same buffer transfer evenly.
Recovery steps
- Re-assemble the sandwich under transfer buffer; roll out bubbles firmlyBubbles are the most common local void pattern.
- Make fresh transfer buffer; verify pH 8.3 and methanol % for PVDF vs nitrocelluloseWrong buffer composition causes edge/centre gradients.
- Extend wet transfer (e.g. 30 V / 90 min → 100 min) for targets >100 kDaLarge proteins often need more time; confirm with Ponceau + residual gel stain.
Controls first
- Does Ponceau on the positive-control / ladder region also look patchy or voided?If yes: Transfer assembly or buffer/power — not antibody detection.
Related symptoms