PCR / Gel: unexpected band size
Amplicon runs at the wrong molecular weight versus the expected/ladder reference.
Ranked causes
- Splice variant or isoform amplified (RT-PCR)Size difference matches a known exon in size.
- Mispriming at a secondary binding siteSingle sharp band, just at the wrong size — not a smear.
- Genomic DNA contamination spanning an intronBand is larger than expected by roughly one intron length.
- Wrong primer pair loaded (tube labels vs plate map)Correlates with a specific well or plate row.
Recovery steps
- Sequence or digest the product to confirm identityFastest way to distinguish mispriming from a real isoform.
- BLAST both primers against the reference genome for secondary sitesIdentifies mispriming before repeating the run.
- DNase-treat template and re-run to rule out genomic contaminationIf the band shrinks to expected size, gDNA was the cause.
- Re-verify primer tube labels against the plate mapRules out a labeling mismatch at setup.
Controls first
- Did the positive control band run at the expected size on the ladder?If no: Ladder, control template, or primer identity issue — check materials before sample interpretation.
- Is the band a single sharp line, not smeared?If yes: Specific amplification of the wrong target — check primer specificity and template identity, not PCR chemistry.